X-Rite - Optronik manufactures spheres with diameters of
50 cm and 100 cm:
Integrating sphere ITS10-50 (50 cm diameter) - General
characteristics
- Equipment for luminous flux measurement of light sources
by comparison with a standard lamp
- Construction according to DIN 5032 section 1 and CIE Publication
No. 84 "The Measurement of Luminous Flux"
- Metal frame construction with swiveling mechanism to open
the sphere
- Light-tight mechanical lock
- Flexible adjustment bar to connect X-Rite Optronik lamp
holders
- 3 burning positions (2 vertical, 1 horizontal), clamping
device for fixing holders in pendent, upright, and horizontal
position
- Coated inside with Optronik BASO4 photometer paint with
Lambertian white diffusing surface, reflectance 80%
with variation of reflectance < 1.5% in wavelength
range 400 to 700 nm
- Diameter: 0.50 meter
- Sphere material: Aluminum
- Sphere paint (outside): RAL 5018
- Thermometer: Digital, resolution 0.1°C
- Universal flange prepared to install several photometer
heads for V (λ) ( Digilumen
9500)
- Baffle to be shifted
- Auxiliary lamp
Integrating sphere ITS10-100 (100 cm diameter) - General
characteristics
- Equipment for luminous flux measurement of light sources
by comparison with a standard lamp
- Construction according to DIN 5032 section 1 and CIE Publication
No. 84 "The Measurement of Luminous Flux"
- Metal frame construction with swiveling mechanism to open
the sphere
- Light-tight mechanical lock
- Flexible adjustment bar to connect X-Rite Optronik lamp
holders
- 3 burning positions (2 vertical, 1 horizontal), clamping
device for fixing holders in pendent, upright, and horizontal
position
- Coated inside with Optronik BASO4 photometer paint with
Lambertian white diffusing surface, reflectance 80%
with variation of reflectance < 1.5% in wavelength
range 400 to 700 nm
- Diameter: 1 meter
- Sphere material: Aluminum
- Sphere paint (outside): RAL 5018
- Thermometer: Digital, resolution 0.1°C
- Universal flange prepared to install several photometer
heads for V(λ) ( Digilumen
9500) and
tristimulus colorimeter head (FMS 10)
- Baffle to be shifted
-Auxiliary lamp
Scope of supply
- Sphere ITS10-50 or ITS10-100
- Stable stand
- Auxiliary lamp
- Baffle
- Thermometer
- Adjustment bar for positioning sample
- Instruction manual
Other sphere diameters
In cooperation with our sister company, Labsphere,
we can offer any kind of lamp measurement system with spheres
as large as 5 meters as well as very small spheres with
a diameter of a few inches.
|